Author: Shenzhen Zhongjing Global Purification Technology Co., Ltd.Time:2018-07-26 11:14:45Views:2991

Medical consumables purification workshop
  • Medical consumables purification workshop

Medical equipment is widely used in A, disposable sterile medical equipment 1, disposable sterile syringe; 2, disposable infusion set; 3, disposable blood transfusion device; 4, disposable anesthesia puncture package; Single-use intravenous infusion needle; 6, disposable sterile needle; 7, disposable plastic blood bag; 8, disposable blood collection device; 9, disposable titration tube infusion device.


requirements

1. Sterile medical devices or single-packaged parts that are implanted into human tissues, directly or indirectly connected to blood, marrow cavity or unnatural cavity, processing (not cleaning) parts, final cleaning, assembly The production area such as the initial packaging and its sealing should be no less than the 100,000 level of cleanliness.

2, YY 0033-2000 sterile medical device production management specifications; 2, YY / T 0567.1-2005 aseptic processing of medical products 3, YY / T 0567.2-2005 aseptic processing of medical products 4, GB 50457-2008 Pharmaceutical industry clean factory design specification

Third, the requirements of temperature and humidity 1, with the requirements of the production process. 2. When there is no special requirement in the production process, the clean room (area) with air cleanliness of 100,000 grades should be 18 °C ~ 26 °C, the relative humidity should be 45% ~ 65%; the air cleanliness is 100,000, thirty Class 10,000 clean room (zone) temperature should be 18 ° C ~ 26 ° C, relative humidity should be 45% ~ 65%. When there are special requirements, it should be determined according to the process requirements. 3. The temperature of the personnel purification room should be 16 °C ~ 20 °C in winter and 26 °C ~ 26 °C in summer. Including: partition wall, ceiling, floor, purification, air conditioning, fire protection, pure water pipeline, pressure difference.

Fourth, the test content: temperature and humidity, cleanliness, pressure difference, gas exchange times, sedimentation bacteria